Disconnecting The Cylinder From The Compressor - Invacare IOH200AW Operator's Manual

Homefill ii compressor
Hide thumbs Also See for IOH200AW:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Disconnecting the Cylinder from the Compressor

NEVER use any type of tool to connect/disconnect the
cylinder and the compressor. Otherwise, severe injury
and/or damage may occur.
NOTE: For this procedure, refer to FIGURE 4.3 and FIGURE 4.4
on page 26.
1. Press the compressor power switch to the Off (O) position.
2. Grasp the cylinder/regulator assembly in the area behind
the cylinder gauge.
3. With the other hand, grasp the outer ring (sleeve) of the
connector fillport and push DOWN.
4. Lift up on the cylinder/regulator assembly to remove from
the connector fillport.
Cylinder
Gauge
Compressor
Connector
Fillport
FIGURE 4.3 Disconnecting the Cylinder from the Compressor
Part No 1145805
WARNING
Cylinder/Regulator
Assembly
Cylinder
Fillport
25
PUSH DOWN
Outer Ring
(Sleeve)
HomeFill
II Compressor
®