SECTION 5—COMPRESSOR
OPERATION
Connecting/Disconnecting the Cylinder
to/from the Compressor
NEVER use tools of any kind to connect/disconnect the cylinder 
and the compressor. Otherwise, severe injury and/or damage 
may occur.

Connecting the Cylinder to the Compressor

DO NOT drop oxygen cylinders. Use two hands when 
handling/transporting oxygen cylinders. Otherwise, injury or 
damage may occur.
DO NOT use a liquid leak detector to test for leaks.
NOTE: For this procedure, refer to FIGURE 5.1 on page 24 and 
FIGURE 5.2 on page 25.
1. Examine the cylinder gauge. If the cylinder pressure is less 
than 1500 psig, proceed to STEP 2.
NOTE: For location of the cylinder gauge, refer to the cylinder's 
manual.
NOTE: If the cylinder pressure is greater than 1500 psig, DO NOT 
attempt to top off this cylinder. It may not fill.
2. Set the regulator flow dial on the cylinder to Off. Refer to the 
cylinder's manual.
3. Remove the cylinder and connector fillport covers (if 
present).
Part No 1100873
SECTION 5—COMPRESSOR OPERATION
WARNING
WARNING
23