Chapter 3—Setting Treatment Parameters
To set an SVS profile:
1) From the "Home"
screen, touch the SVS
Profile button. The SVS
Profile window opens.
3) Touch Start Na+,
and enter the
starting maximum
Na value using the
data entry keypad
on the control panel.
4) Touch Base Na+, and,
using the data entry keypad,
enter the final Na
concentration value to be
reached at the end of the SVS
profile.
Figure 39 – Entering an SVS Program
The SVS timer is activated when the Tx Clock button is initially selected and confirmed to
start treatment. The green light located above the SVS on/off key will illuminate. The SVS
profile parameters can only be changed if the SVS is turned off using the SVS on/off key.
82
2) Select the prescribed profile by touching
the corresponding button from among the
four options located in a row along the top
of the screen.
2008K Hemodialysis Machine Operator's Manual Rev P
5) Touch the SVS Profile
button and, using the data
entry keypad, enter the
running time for the SVS
profile.
6) Press
CONFIRM to save
the SVS settings.