Fresenius Medical Care 2008K Operator's Manual page 230

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Appendix C—The 2008K@home Hemodialysis Machine
Figure 75 – 2008K and 2008K@home "Home" Screens
Table 36 – Software Screen Differences
Item
Setting Treatment
Parameters
Testing the
Machine
Setting Dialysate
Parameters
UF and SVS
Profiles
230
2008K
Set on various screens,
organized by screen-
buttons along bottom of
display screen
All tests are performed on
the "Test & Options"
screen
The "Dialysate" screen is
accessed by touching the
Dialysate screen-button
along the bottom of the
Touchscreen
Available from the
"Home" screen
2008K Hemodialysis Machine Operator's Manual Rev P
2008K@home
Treatment parameters are grouped on
the "Rx Parameter" screen (accessed
from the "Select Program" screen) and
programmed all at once outside of
Dialysis Mode.
The UF goal is calculated based on the
UF Time (called 'Tx Time') and patient
weight entered in the guided "Tx Set up"
screens.
Tests are performed during guided
setup; during treatment, tests are
available on the "Help" screen
Concentrates are selected in the "Rx
Parameters" screen.
Concentrates are confirmed in the
guided set up process. Touching the
Conductivity button on the "Home"
screen will display these settings if limits
need to be adjusted.
The 'In Center' option must be set in
order to display the Dialysis Mode SVS
Profile and UF Profile buttons.
To set the 'In Center' option: in Service
Mode, on the "Options: Treatment
Options" screen, touch the In Center Set
up toggle button to set it to 'In Center'
then press the CONFIRM key.