160
E.04 (Arterial or SN
Reference Voltage error
Blood Pump Message)
E.05 (Arterial or SN
EEPROM error
Blood Pump Message)
E.06 (Arterial or SN
Watchdog timeout
Blood Pump Message)
E.07 (Arterial or SN
+ 12 volts is outside the allowable
Blood Pump Message)
range
E.08 (Arterial or SN
+ 24 volts is outside the allowable
Blood Pump Message)
range
E.09 (Arterial or SN
- 12 volts is outside the allowable
Blood Pump Message)
range
E.10 (Arterial or SN
+ 5 volts is outside the allowable
Blood Pump Message)
range
E.14 (Arterial or SN
50 ms second time period exceeded
Blood Pump Message)
E.15 (Arterial or SN
Software task was not completed
Blood Pump Message)
correctly
E.23 (Arterial or SN
Pump rotor turning when it should not
Blood Pump Message)
be for a second time
E.97 (Arterial or SN
Error copying data into Flash ROM
Blood Pump Message)
while in Service Mode
E.98 (Arterial or SN
Error erasing Flash ROM while in
Blood Pump Message)
Service Mode
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.