Most dialysate or dialysate-related alarm parameters are accessed from the "Dialysate"
screen. Unless otherwise described, enter or change a dialysate-related value by following
the procedure described in "Entering a Parameter" on page 59.
What to do from this screen...
Enter the prescribed dialysate settings for:
Concentrate type
Base Na
Bicarbonate level
Sodium Variation (SVS) profile
Set Alarm limits for:
Low Acid/Bicarbonate alert
Position and width of Conductivity Alarm window
Warning! The specific concentrate and sodium and bicarbonate settings must be prescribed
by a physician.
Note: If the machine is set up for use with Citrasate®, a 'Citrate' meter box will be
displayed in the dialysate constituent list (Functional board software version 3.36 or later).
Acid Concentrate Name
Theoretical
Conductivity (TCD)
of the Final
Dialysate at the
displayed Base Na
and Bicarbonate
values.
Electrolyte Profile of
Selected
Concentrate at the
base sodium and
bicarbonate levels
Figure 28 – The Dialysate Screen
2008K Hemodialysis Machine Operator's Manual Rev P
+
level
Chapter 3—Setting Treatment Parameters
Acid Concentrate Type
Alarm Window
Conductivity
Scale
Actual Conductivity
61