Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
Fail * Neg Flow On
Negative flow on pressure holding
test failed
Fail * Neg Stabilize
Negative flow stabilize test failed.
Fail * Optical Detect
Optical Detector test has failed.
Fail * Pos Flow Off
Positive flow off pressure holding test
failed
Fail * Pos Stabilize
Positive flow stabilize test failed.
Fail * (Remove Air)
Remove air test failed.
Failed Sending Data to
Functional to Actuator board
Actuator Board
communication problem during
startup.
Fail * Temp High Soft
Temperature test has failed.
Fail * Temp Low Soft
Fail * Temp High Hard
Temperature test has failed.
Fail * Temp Low Hard
Fail * TMP High Soft
Transmembrane Pressure (TMP) test
has failed.
Fail * TMP Low Soft
Fail * TMP High Hard
Fail * TMP Low Hard
163
Type
Action Required
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test, if failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Test
Verify stable temp. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician to recalibrate.
Test
Verify stable temp. Rerun test. If failure repeats, remove machine
Message
from service and alert a qualified service technician to recalibrate.
Test
Rerun test. If failure repeats, remove machine from service and alert
Message
a qualified service technician.
2008K Hemodialysis Machine Operator's Manual Rev P