1) The electrolyte profile of the
highlighted selection is
displayed in this column.
Note: If the machine is set up
for use with Citrasate®, a
'Citrate' meter box will be
displayed in the dialysate
constituent list.
2) Highlight the desired
selection by scrolling through
the choices using the or
(up or down) keys on the data
entry keypad. Acetate
formulations can also be
selected as a treatment option.
3) When the desired type is highlighted, press
the CONFIRM key on the control panel to save
the selection and return to the "Dialysate" screen.
Note: Be sure the acid/bicarbonate concentrates
connected to the machine match the type
selected from the Concentrate menu.
4) Enter the prescribed Base Na+
and Bicarbonate values as
described in "Setting Treatment
Parameters."
Note: The Bicarbonate button
displays the available bicarbonate
in the Final Dialysate (post-
reaction of the acid and
bicarbonate concentrates).
Figure 30 – Entering Concentrate Information
2008K Hemodialysis Machine Operator's Manual Rev P
Chapter 3—Setting Treatment Parameters
67