Chapter 6—Alarms and Troubleshooting
Message
Purpose of Message
E.99 (Arterial or SN
Transmit error during Flash update
Blood Pump Message)
while in Service Mode
EEPROM already used,
Advisory message when uploading
Power Off, Replace
hardware key option
EEPROM
EEPROM Missing or
During startup, the machine cannot
Reading Error
properly read the EEPROM memory
chip
Emptying
The blue dialysate line connector is
on the shunt with door closed, the red
dialysate line connector remains on
the dialyzer in order to drain the
dialysate compartment.
Emptying Stopped
When air is sensed, emptying will
stop.
Emptying too long
The dialyzer empty program has
exceeded its maximum limit.
Error Reading Flash
Electronic Self Test
Fail * 9 Volt Battery
9V Power Failure Battery test has
failed.
Fail *Actuator Arterial
Arterial Pressure test has failed.
High
Fail *Actuator Arterial
Low
Fail * Actuator
Conductivity test has failed.
Conductivity High
Fail * Actuator
Conductivity Low
161
Type
Action Required
Alarm
If this message occurs in dialysis mode, return the blood to the
patient if during a treatment. Take the blood pump module out of
service and alert a qualified service technician.
Advisory
Put in a new hardware key or calibration EEPROM in IC 20 and
power up.
Opening
Turn the machine off and try to power up again. If the message
Screen
repeats, take the machine out of service and alert a qualified service
Message
technician.
Warning
If this message occurs when the dialyzer is not being emptied, take
the machine out of service and alert a qualified service technician.
Warning
Connect the red dialyzer return line to the shunt. If the warning is
repeated, take the machine out of service and alert a qualified service
technician.
Alarm
If blood is not sensed, return the dialyzer supply and return lines to
the shunt and close the shunt door to terminate the program. If the
machine was in dialysis (blood sensed), turn machine power off and
back on to clear the program.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Test
Replace Battery
Message
Test
Rerun test, if failure repeats, remove from service and alert a
Message
qualified service technician.
Test
Rerun test, if failure repeats, remove from service and alert a
Message
qualified service technician.