Must Run Test First
The Forced Test is required before
proceeding with UF or SVS.
Need Blood Sensed
An action has been initiated that
requires that blood is sensed.
Neg. Access Flow value
This message is an advisory
message that the Access Flow test
result was a negative value. A
positive value is expected.
Negative error AF value
This message is an advisory
message that the Access Flow test
result was an erroneous value
New Art Limits chosen
This message advises the operator
that a new set of arterial limits has
been set.
New features loaded,
Advisory message when uploading
Power Off, Replace
hardware key option
EEPROM
New TMP Limits Chosen
This message confirms that a new set
of TMP limits have been set.
New Venous Limits?
This message is a prompt for the
operator to set new venous alarm
limits.
New Ven Limits Chosen
New venous alarm limits are set
No Air Detector Alarm
The Prime key has been pressed. A
level detector alarm must exist for this
function to occur
Chapter 6—Alarms and Troubleshooting
2008K Hemodialysis Machine Operator's Manual Rev P
Warning
To proceed with the selected operation, run the Pressure and Alarm
tests. For more information, see Chapter 2 "Testing the 2008K
hemodialysis machine."
Verify venous bloodline is in the Optical Detector.
Dialog
Message
Warning
This can occur if the bloodlines were initially connected in the
reversed position. Press CONFIRM to clear the message. Check that
the bloodlines are properly connected and repeat the Access Flow
test.
Warning
Press CONFIRM to clear the message. Check that the bloodlines are
properly connected and repeat the Access Flow test. Press
CONFIRM to clear the message.
Advisory
Advisory only. No action is required.
Advisory
Put the original calibration EEPROM in IC 20 and restart the
machine.
Advisory
Advisory only. No action is required.
Advisory
To set new venous alarm limits, press RESET and hold.
Advisory
Advisory only. No action is required
Advisory
If the venous chamber has fluid detected, the prime function will not
occur. Press RESET to start the blood pump. If a level detector
alarm occurs, then press the Prime key.