1-16. Procedure for Measuring Luminance

1-16-1
In the Case of Insuffi cient Luminance
Required equipment/tool
. Laser distance meter
. Illuminance meter
. Luminance meter (CL-200)
. String (capable of measuring the length longer than the screen width)
. SRX Controller (CSC adjustment)
. 3D glasses (in the case of 3D)
Procedure
When the luminance is insuffi cient, perform the check in the following procedure.
1. Check the lamp bulb.
There should be no clearance between the end face of the cathode terminal and the mounting face of
the lamp bayonet cap (portion A).
Lamp bayonet cap
2. Check the lamp input.
Check that the lamp wattage matches the loaded lamp in the LAMP TIMER RESET menu in MAIN-
TENANCE of the SRX Controller.
n
In this unit, when the lamp of 4.2 kw is loaded, "4.0 kw" is displayed.
3. Check the lamp optical axis adjustment.
Adjust the lamp optical axis in the order of Z axis → X and Y axes → Z axis. At that time, make sure
that the LCD value decreases at around each maximum point of X, Y and Z axes, and adjust each to
the maximum value. When the LCD value does not decrease at around the maximum value, there
may be a failure in the loading of lamp, or a failure in the equipment.
4. Perform the CSC adjustment. (Refer to Installation Manual.)
Before adjusting CSC, be sure to perform a reset.
n
When you repeat the CSC adjustment without performing a reset, the luminance may be decreased.
1-76 (E)
Cathode terminal
Portion A
SRX-R320