Troubleshooting - ResMed escape s8 User Manual

Resmed mask user's manual
Hide thumbs Also See for escape s8:
Table of Contents

Advertisement

Available languages

Available languages

T
ROUBLESHOOTING
If there is a problem, try the following suggestions. If the problem cannot be solved,
contact your equipment supplier or ResMed. Do not attempt to open the device.
Problem
No display.
Insufficient air
delivered from
S8 Escape.
Excessive air delivered
from the S8 Escape.
26
Possible Cause
Power not connected.
Ramp Time is in use.
Air filter is dirty.
Air tubing is kinked or
punctured.
Air tubing not connected
properly.
Mask and headgear not
positioned correctly.
Cap(s) missing from access
port(s) on mask.
Altitude compensation setting is
incorrect.
Pressure required for treatment
may have changed.
Altitude compensation setting is
incorrect.
Solution
Ensure the appropriate power
cord is connected to your
S8 Escape and the power switch
(if available) is on.
Wait for air pressure to build up
or change ramp time.
Replace air filter.
Straighten or replace tubing.
Check air tubing.
Adjust position of mask and
headgear.
Replace cap(s).
Adjust altitude compensation
setting to the correct altitude
range.
See your clinician to adjust the
pressure.
Adjust altitude compensation
setting to the correct altitude
range.