Probable Causes
Suggested Actions
Troubleshooting and Test Functions
Valve Fuse
Valve Fuse
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent-selection valves in the binary pump has drawn excessive
current causing the selection-valve electronic fuse to open.
• Defective solvent selection valve.
• Defective connection cable (front panel to HPM board).
• Defective HPM board.
Restart the binary pump. If the error message appears again, exchange the
solvent selection valve.
Exchange the connection cable.
Exchange the HPM board.
70