Repairing the Pump
Exchanging the Pump Seals and Seal Wear-in Procedure
2 Using one of the plungers carefully remove
the seal from the pump head (be careful, not
to break the plunger). Remove wear
retainers, if still present.
4 Insert seals into the pump head and press
firmly in position.
Seal
122
Plunger
Seal
3 Clean the pump chambers with lint free cloth.
Ensure all particulate matter is removed. Best
cleaning results will be achieved by removing
all valves (see pages 109, 112) and the
capillary. Inject solvent into each chamber.
5 Reassemble the pump head assembly (see
"Reassembling the Pump Head Assembly" on
page 129). Reset the seal wear counter and
liquimeter as described in the User Interface
documentation.
Chambers