Installing the Pump
Site Requirements
Never use cables other than the ones supplied by Agilent Technologies
to ensure proper functionality and compliance with safety or EMC
regulations.
Bench Space
The binary pump dimensions and weight (see Table 1) allow to place the
binary pump on almost any laboratory bench. It needs an additional 2.5 cm
(1.0 inches) of space on either side and approximately 8 cm (3.1 inches) in
the rear for the circulation of air and electric connections.
If the bench should carry a complete Agilent 1100 Series system, make sure
that the bench is designed to carry the weight of all the modules.
The pump should be operated in a horizontal position!
Environment
Your binary pump will work within specifications at ambient temperatures
and relative humidity as described in Table 1.
Do not store, ship or use your binary pump under conditions where
temperature fluctuations could cause condensation within the binary pump.
Condensation will damage the system electronics. If your binary pump was
shipped in cold weather, leave it in its box and allow it to warm slowly to room
temperature to avoid condensation.