Repairing the Pump
Exchanging a Pump Drive
6 Place the new pump drive into the recess in
the foam part and connect the cables to the
connectors on the main board (J16, J17 or
J12, J28).
Cables
8 Reinstall the pump heads and secure with the
two screws.
Screws
150
7 Replace the damper and the Z-panel and
secure it with the two screws. Reconnect the
cables to the main board (J19, J28).
Screw
9 Reconnect all capillaries and cables. Secure
the mixing chamber in its holder.
Damper
Active inlet valves
Damper
Z-panel
Screw
Mixing
chamber