Figure 8-4
Load
L=0
L0
L1
L2
L
max
L
max
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02
Linearizing the scales' characteristic curve
Comment
100 kg load cell (2 mV/V) not
loaded
Mechanical installation on load cell
(dead load)
Calibration weight 1 placed onto
scale
Calibration weight 2 placed onto
scale
Nominal weight of load cell
+10 %
Rated weight + approx. 10 %
Scale parameters and functions
8.4 Calibration procedure
Load
e.g. 25 kg
e.g. 60 kg
e.g. 80 kg
100 kg
Approx. 110 kg
Digits
Approx. 0
Approx. 500 000
Approx. 1 200 000
Approx. 1 650 000
Approx. 2 000 000
Approx. 2 200 000