Table 13- 14 Requirements: Interference Immunity In Industrial Area In Accordance With En 61000-6-2 - Siemens SIWAREX WP231 Manual

Hide thumbs Also See for SIWAREX WP231:
Table of Contents

Advertisement

Table 13- 14 Requirements: Interference immunity in industrial area in accordance with EN 61000-6-2

Comments
Burst pulses on power supply cables
Burst pulses on data and signal cables
Electrostatic discharge (ESD)
Electrostatic air discharge (ESD)
Surge on power supply cables
Surge on data and signal cables
HF irradiation amplitude modulated
HF irradiation, cell phone frequencies
HF voltage on data, signal and power
supply cables 0.15 to 80 MHz
Not applicable for shielded cables and symmetrical ports
1)
* An external protection element has to be installed to meet the requirement (e.g.: Blitzductor VT
AD24V, Dehn&Söhne)
NOTICE
Radio interference is possible
This is a device of class A. The device may cause radio interference in residential areas.
Implement appropriate measures (e.g.: use in 8MC cabinets) to prevent radio interference.
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02
13.2 Electrical, EMC and climatic requirements
Standard
Severity class
EN45501
1 kV
OIML R 76
EN 61000-4-4
2 kV
NAMUR NE21
EN 61326
EN 61000-4-2
6 kV direct/indirect
NAMUR NE21
EN 61326
EN 45501
OIML R 76
EN 61000-4-2
8 kV
NAMUR NE21
EN 61326
EN 45501
OIML R 76
EN 61000-4-5
1 kV symmetrical
IEC 61131-2
2 kV asymmetrical
NAMUR NE21
EN 61326
EN 61000-4-5
1 kV symmetrical
IEC 61131-2
2 kV asymmetrical
NAMUR NE21
EN 61326
IEC61000-4-3
80 to 2 000 MHz: 12 V/m
NAMUR NE21
Mod.: 80% AM with 1 kHz
OIML R76
EN 45501*3
Note:
In the ranges 87 ... 108 MHz, 174 ... 230 MHz
and 470 ... 790 MHz: 3 V/m
IEC 61000-4-3
900 MHz (± 5 MHz)
1.89 Ghz (± 10 MHz)
10 V/m
IEC 61000-4-6
10 kHz to 80 MHz: 10 Veff
NAMUR NE21
Mod.: 80% AM with 1 kHz
EN 61326
OIML R 76
Technical data
1)
171

Advertisement

Table of Contents
loading

Table of Contents