Y1152A Switch Control

The switch control for the Y1152A distribution board is shown below. All
channels are driven in the PAIRed mode (Preset state). See Figure 4-12 for
additional information. See Chapter 2 for bank-specific channel numbering.
* Note that for switches connected to SW1, path closed is accomplished with
the ROUTe:OPEN command and path opened is accomplished through the
ROUTE:CLOSe command.
Agilent 34945A Microwave Switch/Attenuator Driver Module User's Guide
Channel / Switch 2
Channel 1xx1 / SW2 Path 1
Channel 1xx2 /SW2 Path 2
Channel 1xx3 /SW2 Path 3
Channel 1xx4 /SW2 Path 4
Channel 1xx5 /SW2 Path 5
Channel 1xx6 /SW2 Path 6
Channel 1xx7 /SW2 Open All 1
Channel 1xx8 /SW2 Open All 2
Channel / Switch
Channel 1xx1 /SW1 Path1
Channel 1xx2 /SW1 Path2
Channel 1xx3 /SW1 Path3
Channel 1xx4 /SW1 Path4
Channel 1xx5 /SW1 Path5
Channel 1xx6 /SW1 Path6
Channel / Switch
Channel 1xx7 /SW2
Channel 1xx8 /SW3
Additional Operating Information
Path Closed
ROUT:CLOS (@1xx1)
Close another path or open all
ROUT:CLOS (@1xx2)
Close another path or open all
ROUT:CLOS (@1xx3)
Close another path or open all
ROUT:CLOS (@1xx4)
Close another path or open all
ROUT:CLOS (@1xx5)
Close another path or open all
ROUT:CLOS (@1xx6)
Close another path or open all
ROUT:CLOS (@1xx7)
note: use either channel
ROUT:CLOS (@1xx8)
Path State
Path Closed*
ROUT:OPEN (@1xx1)
ROUT:OPEN (@1xx2)
ROUT:OPEN (@1xx3)
ROUT:OPEN (@1xx4)
ROUT:OPEN (@1xx5)
ROUT:OPEN (@1xx6)
Switch State
State A
ROUT:OPEN (@1xx7)
ROUT:OPEN (@1xx8)
Path Open
to open all
Path Opened*
ROUT:CLOS (@1xx1)
ROUT:CLOS (@1xx2)
ROUT:CLOS (@1xx3)
ROUT:CLOS (@1xx4)
ROUT:CLOS (@1xx5)
ROUT:CLOS (@1xx6)
State B
ROUT:CLOS (@1xx7)
ROUT:CLOS (@1xx8)