Switch Control; Figure 2-1 Switch Matrix Concept - Agilent Technologies U2751A User's And Service Manual

Modular switch matrix
Table of Contents

Advertisement

Switch Control

U2751A User's and Service Guide
A matrix switch connects multiple inputs to multiple
outputs. A matrix is arranged in rows and columns. For
example, the U2751A is a 4×8 matrix that can be used to
connect four sources to eight test points as shown in
Figure 2-
1.
Any column can be connected to any row by activating the
corresponding relay that connects the column to the row as
shown in
Figure 2-
has its own unique channel label representing the row and
column. For example, channel 302 represents the cross-point
connection between row 3 and column 2.
Be aware that it is possible to connect more than one source
to the same point with a matrix. It is vital to make sure that
these connections do not create dangerous or unwanted
conditions.

Figure 2-1 Switch matrix concept

Operation and Features
1. Each cross-point relay on this module
Note: Three-digit channel numbers are derived from
the intersection of the rows and columns, with the
last two digits representing the columns. The
intersection shown represents channel 202
(Row 2, Column 2)