Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
Internal
0x1120 highPressureAlarm
Fault
0x1121 lowPressureAlarm
Internal
Fault
Internal
0x1122 compressorRunningAlarm
Fault
0x1123 compressorNotRun-
Internal
ningAlarm
Fault
36
Display Message
Possible Reason
The display shows:
1. Blockage in outlet
"Shut Down Internal
Fault"
The display shows:
1. Internal leakage of
"Shut Down Internal
Fault"
2. Blockage of inlet
The display shows:
1. Compressor or
"Shut Down Internal
Fault"
2. The compressor
1. Compressor or
The display shows:
"Shut Down Internal
Fault"
2. The compressor
Solution
1. Remove any kinks
gas stream.
or blockage in
Blockage may
cannula.
be internal or
2. Replace the
external.
patient outlet filter.
3. If alarm persists,
contact Invacare
for service.
1. Clean air intake
pneumatic circuit.
filter screen.
2. If alarm persists,
flow.
contact Invacare
for service.
1. If alarm persists,
compressor's
contact Invacare
wire harness is
for service.
damaged. -or-
wire harness is
damaged or loose.
1. If alarm persists,
compressor's
contact Invacare
wire harness is
for service.
damaged. -or-
wire harness is
damaged or loose.