Invacare POC1–100B-C Service Manual page 31

Platinum mobile
Hide thumbs Also See for POC1–100B-C:
Table of Contents

Advertisement

General
Alarm
Alarm Code Name
Alarm
Code
Name
Battery Too
0x1102 SDA_batteryTooCold
Cold (shut
down)
Unit Too
0x1103 SDA_unitTooHot
Hot (shut
down)
Unit Too
0x1104 SDA_unitTooCold
Cold (shut
down)
1187375-B
Display Message
Possible Reason
The display shows:
1. The concentrator is
"Shut Down Battery
Too Cold"
The display shows:
1. The concentrator is
"Shut Down Unit Too
Hot"
The display shows:
1. The concentrator is
"Shut Down Unit Too
Cold"
Solution
1. Remove the cold
running off battery
battery and allow
power and at least
to warm.
one of the battery
2. Change the
temperatures is
battery.
less than or equal
3. If alarm persists,
to
contact Invacare
–20° C.
for service.
1. Allow unit to cool.
running off battery
2. Clean intake filter
power or external
screen.
power and the
3. If alarm persists,
unit temperature
contact Invacare
is greater than or
for service.
equal to 55°C.
1. Allow unit to
running off battery
warm.
power or external
2. If alarm persists,
power and the unit
contact Invacare
temperature is less
for service.
than or equal to 0°
C.
Troubleshooting
31