Invacare POC1–100B-C Service Manual page 27

Platinum mobile
Hide thumbs Also See for POC1–100B-C:
Table of Contents

Advertisement

General
Alarm
Alarm Code Name
Alarm
Code
Name
Battery
0x6206 LPA_batteryCold
Cold
(Running)
Unit Hot
0x5203 LPA_unitHot
(Running)
0x5204 LPA_unitCold
Unit Cold
(Running)
1187375-B
Display Message
Possible Reason
The display shows:
1. The concentrator is
"Battery Cold Allow to
Warm"
The display shows:
1. The concentrator is
"Unit Hot Allow to
Cool"
The display shows:
1. The concentrator is
"Unit Cold Allow to
Warm"
Solution
1. Remove the cold
running off battery
battery and allow
power and at least
to warm.
one of the battery
2. If alarm persists,
temperatures is
contact Invacare
less than or equal
for service.
to –15° C
1. Allow the unit to
running off battery
cool.
power or external
2. Clean air intake
power (AC/DC) and
filter.
the unit internal
3. If alarm persists,
temperature is
contact Invacare
greater than or
for service.
equal to 50° C.
1. Allow unit to
running off battery
warm.
power or external
2. If alarm persists,
power (AC/DC) and
contact Invacare
the unit internal
for service.
temperature is less
than or equal to
5°C.
Troubleshooting
27