Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
Battery
0xb20
LPA_batteryTooHot-
Too Hot
b
ToCharge
To Charge
(Running)
Battery
0xb20c
LPA_batteryTooCold-
Too Cold
ToCharge
To Charge
(Running)
Battery Hot
0x6205 LPA_batteryHot
(Running)
26
Display Message
Possible Reason
The display shows:
1. The concentrator
"Cannot Charge Allow
to Cool"
The display shows:
1. The concentrator
"Cannot Charge Allow
to Warm
The display shows:
1. The concentrator is
"Battery Hot Allow to
Cool"
Solution
1. Remove the hot
is running off
battery and allow
external power
to cool.
(AC/DC) and
2. If alarm persists,
at least one
contact Invacare
of the battery
for service.
temperatures is
greater than or
equal to 49° C.
1. Remove the cold
is running off
battery and allow
external power
to warm.
(AC/DC) and
2. If alarm persists,
at least one
contact Invacare
of the battery
for service.
temperatures is
less than or equal
to 8° C.
1. Remove the hot
running off battery
battery and allow
power and at least
to cool.
one of the battery
2. If alarm persists,
temperatures is
contact Invacare
greater than or
for service.
equal to 55°C.
1187375-B