Invacare POC1–100B-C Service Manual page 33

Platinum mobile
Hide thumbs Also See for POC1–100B-C:
Table of Contents

Advertisement

General
Alarm
Alarm Code Name
Alarm
Code
Name
Stuck
0x1107 SDA_stuckButton
Button
No Breath
0x110c
SDA_noBreathDetected
Detected
(shut down)
0x1109 ricEventQueueFullAlarm
Internal
Fault
1187375-B
Display Message
Possible Reason
The display shows:
1. One of the
"Shut Down Stuck
Button"
The display shows:
1. An inhalation has
"Shut Down No Breath
Detected"
The display shows:
1. PCB or software
"Shut Down Internal
Fault"
Solution
1. Make sure no
buttons on the
control panel
concentrator's
buttons are
control panel was
pressed.
pressed when a
2. Disconnect all
power source is
power sources
being connected
then reconnect to
to the unit.
a power source.
3. If alarm persists,
contact Invacare
for service.
1. Verfy cannula is
not been detected
connected, not
by the unit for
kinked, properly
more than 120
positioned and
seconds.
user is breathing
through his/her
nose.
2. Check the outlet
filter/Replace filter.
3. If alarm persists,
contact Invacare
for service.
1. If alarm persists,
fault.
contact Invacare
for service.
Troubleshooting
33