Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
No Breath
0x3201 LPA_noBreathDetected
Detected
(Running)
0x4202 LPA_breathRateTooHigh
Breath Rate
Too High
(Running)
24
Display Message
Possible Reason
The display shows:
1. The concentrator
"No Breath Detected
Check Cannula" (not
displayed if in "Warm
Up" or "Prolonged
Warm Up")
2. Restriction after
The display shows:
1. The user breath
"Breath Rate High
Reduce Activity"
Solution
1. Verify cannula is
has not detected
connected, not
an inhalation for
kinked, properly
more than 15
positioned and
seconds. -or-
user is breathing
through his/her
the conserver
nose.
valve.
2. Check the outlet
filter/Replace filter.
3. If alarm persists,
contact Invacare
for service.
1. Make sure breath
rate has exceeded
rate is lower than
the capacity of
40BPM.
the concentrator
2. If alarm persists,
for more than 15
contact Invacare
seconds.
for service.