Invacare® Platinum® Mobile
General
Alarm
Alarm Code Name
Alarm
Code
Name
Depleted
0x1100 SDA_depletedBattery
Battery
Battery Too
0x1101 SDA_batteryTooHot
Hot (shut
down)
30
Display Message
Possible Reason
The display shows:
1. The concentrator is
"Shut Down Battery
Depleted"
The display shows:
1. The concentrator is
"Shut Down Battery
Too Hot"
Solution
1. Supply unit with
running off battery
external power
power and the
(AC/DC) or install
battery charge
a charged battery
level is less than
into the unit.
or equal to 5%.
2. If alarm persists:
a. Inspect
b. Reinstall
3. If alarm persists,
contact Invacare
for service.
1. Remove the hot
running off battery
battery and allow
power and at least
to cool.
one of the battery
2. Change the
temperatures is
battery.
greater than or
3. If alarm persists,
equal to 60°C.
contact Invacare
for service.
the battery
connector for
debris.
battery.