Invacare® TDX® SP2 Series
located under the seat.
Center mount powered legrests and tilt module
G-Trac® Sensor
The optional G-Trac-sensor is located behind the rear shroud.

6.3 Chassis

6.3.1 Stability lock

Replacing gas cylinder
CAUTION!
Risk of crushing
Mobility device is very heavy. Injury hazard to
hands and feet.
– Use proper lifting techniques.
Risk of injury by uncontrolled movement of
mobility device
– Switch power supply off (ON/OFF key).
– Engage drive.
– Before raising mobility device, secure wheels by
blocking them with wedges.
3/16" Allen key
5/16" wrench with extension
3/4" wrench
5/32" Allen key
7/16" socket wrench
Phillips screwdriver size 2
Torque wrench 5 - 25 Nm (or similar)
Two long wooden blocks, min. 14 x 14 x 30 cm
Thread locking adhesive, for example Loctite 242
Removing gas cylinder
1. Lift mobility device up on one side and place a 14 cm
high wooden block under battery case so that wheels
are off the ground and can rotate freely. Use proper
lifting techniques.
2. Repeat same procedure on other side of mobility device.
14
3.
Loosen and remove hand screw A under rear light B
near wheel D.
4.
If mobility device is fitted with cable ties, remove.
When installing, make sure to replace cable ties.
Loosen and remove screw A.
5.
Loosen and remove screw A over mud guard B under
seat frame C.
6. Remove batteries as described in 6.7.6 Replacing
batteries, page 33 and 6.7.9 Replacing batteries/replacing
battery cables, page 35.
7. Remove drive wheel as described in 6.5.6 Replacing
drive wheel, page 26.
Fig. 6-4
Fig. 6-5
1532486-L