Problem
Motors do not run
Motor makes clicking
noise
Scraping noise or
motor blocked
Gearbox makes clicking
noise
44
Other symptoms
Possible cause
None
Bad connections
Fuse burnt out
Batteries defective
Cabling to power
module or remote
defective
Power module
defective
Corroded contacts
Contacts defective
None
Clutch(es) defective
Bearing defective
Collector defective
None
Clutch(es) defective
Bearing defective
Gearbox defective
None
Gearbox defective
Drive wheel loose
Documentation
Solution
Check all cabling
See 6.7.15 Checking
cable, page 38.
Check cabling and
See 6.7.15 Checking
replace fuse
cable, page 38 and
6.7.13 Checking and
replacing main fuse,
page 37.
Replace batteries
See 6.7.6 Replacing
batteries, page 33
and6.7.9 Replacing
batteries/replacing
battery cables, page 35.
Check cabling
See 6.7.15 Checking
cable, page 38.
Replace power module
See 6.7.1 Replacing
power module, page 31.
Check cabling, replace
See 6.7.15 Checking
if necessary
cable, page 38.
Replace clutch
See 6.4.7 Replacing
motor/gearbox unit,
page 23.
Replace motor
See 6.4.7 Replacing
motor/gearbox unit,
page 23.
See 5.1 Testing
Measure internal
resistance of motor,
motor, page 11
replace motor if
and 6.4.7 Replacing
motor/gearbox unit,
defective
page 23
Replace clutch
See 6.4.4 Replacing
motor/gearbox clutch,
page 22.
Replace motor
See 6.4.7 Replacing
motor/gearbox
unit, page 23 and
6.4.4 Replacing
motor/gearbox clutch,
page 22.
Replace gearbox
See 6.4.7 Replacing
motor/gearbox
unit, page 23
and6.4.4 Replacing
motor/gearbox clutch,
page 22.
Replace gearbox
See 6.4.7 Replacing
motor/gearbox
unit, page 23 and
6.4.4 Replacing
motor/gearbox clutch,
page 22.
Tighten drive wheel,
See 6.4.7 Replacing
secure bolts with
motor/gearbox unit,
Loctite if necessary
page 23.
1532486-L

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tdx sp2 series

Table of Contents