Requesting An Echo Profile; Device Address; Auto False Echo Suppression - Siemens SITRANS LR260 Operating Instructions Manual

Radar, hart
Hide thumbs Also See for SITRANS LR260:
Table of Contents

Advertisement

Requesting an Echo Profile

a.
In PROGRAM mode, navigate to:
LEVEL METER > DIAGNOSTICS >
ECHO PROFILE (3.1)
b.
Press RIGHT arrow
profile.
c.
In the Profile screen, press UP
arrow
to select the Transmit
icon, and Right ARROW
update the profile.
d.
Press DOWN arrow
the Exit icon, then Right ARROW
to return to previous menu.

Device Address

The default address for the device is 0. To reset the address via the handheld
programmer, see

Auto False Echo Suppression

If SITRANS LR260 displays a false high level, or the reading is fluctuating between the
correct level and a false high level, you can use the Auto False Echo Suppression
parameters to prevent false echo detection. See
setup (TVT SETUP)
Page 34
SITRANS LR260 (HART) – INSTRUCTION MANUAL
to request a
to
to select
Device Address
on page 80.
for instructions.
DIAGNOSTICS
3.1
ECHO PROFILE
MEAS. VALUES
REMAIN. DEV. LIFE
REMAIN. SENS. LIFE
transmit icon,
selected
2.2.5. TVT (Auto False Echo Suppression)
transmit icon,
deselected
exit icon,
selected
exit icon,
deselected
7ML1998KE03

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents