4. Repeat this procedure for all time marker values recommended for your oscilloscope.
5. Remove the signal by pressing .

Summary of Commands and Queries

This section describes commands and queries that are used specifically for the
oscilloscope calibration option. Each command fall into one or more command
categories: Sequential, Overlapped, or Coupled.
Sequential Commands – Commands executed immediately as they are encountered in
the data stream are called sequential commands. For more information, see "Sequential
Commands" in Chapter 5 of the 5080A Operators Manual.
Overlapped Commands – Commands that require additional time to execute are called
overlapped commands because they can overlap the next command before completing
execution. To be sure an overlapped command is not interrupted during execution, use
the *OPC, *OPC?, and *WAI commands to detect command completion. See Table 6-8
for all the commands that are classified as overlapped. For more information, see
"Overlapped Commands" in Chapter 5 of the 5080A Operators Manual.
Coupled Commands – These are called coupled commands (examples: CUR_POST and
OUT) because they "couple" in a compound command sequence. Care must be taken to
be sure the action of one command does not disable the action of a second command and
thereby cause a fault. See Table 6-8 for all the commands that are classified as coupled.
For more information, see "Coupled Commands" in Chapter 5 of the 5080A Operators
Manual.
18
Repeat for digital and analog mode as required. Some oscilloscopes may need the
magnification changed while calibrating in analog mode.
riešenia na presné meranie
http://www.elso.sk
Peaks are aligned
with center axis
gl011i.eps