Agilent Technologies 1260 Infinity User Manual page 117

Nanoflow pump
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

8
Test Functions and Calibration
Leak Test
Plateau 2
Piston B2 delivers with a flow rate of 3 µL/min for 30 s.
Ramp 5
Piston A1 delivers 50 µL/min for approximately 8 s.
Plateau 3
Piston A1 delivers with a flow rate of 3 µL/min for 30 s.
Ramp 6
Piston B1 delivers 50 µL/min for approximately 7 s.
Plateau 4
Piston B1 delivers with a flow rate of 3 µL/min for approximately 30 s. At the
end of the fourth plateau, the test is finished and the pump switches off.
1260 Nanoflow Pump User Manual