Endress+Hauser Cerabar S PMP7 Operating Instructions Manual
Endress+Hauser Cerabar S PMP7 Operating Instructions Manual

Endress+Hauser Cerabar S PMP7 Operating Instructions Manual

Process pressure measurement with mid part certificate

Advertisement

Quick Links

Operating Instructions – Description of Device Functions
Cerabar S PMP71 with MID Part Certificate
Process pressure measurement
BA00413P/00/EN/03.14
71270370
Valid for software version:
02.10.54

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Cerabar S PMP7

  • Page 1 Operating Instructions – Description of Device Functions Cerabar S PMP71 with MID Part Certificate Process pressure measurement BA00413P/00/EN/03.14 71270370 Valid for software version: 02.10.54...
  • Page 2: Serial Number

    Cerabar S 4 to 20mA HART Order code 00X00-XXXX0XX0XXX Ser. No.: X000X000000 TAG No.: XXX000 Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3: Table Of Contents

    ......47 Index ......54 Endress+Hauser...
  • Page 4: Notes On Use

    In Section 5, all of the parameters in a function group are summarized in a table. Finding parameter description using parameter names (index) The index lists all the parameters in alphabetical order. The page reference/link takes you to the parameter in question. Endress+Hauser...
  • Page 5: Finding Parameter Description Using Id

    CONFIG RECORDER DEVICE SERIAL No PCB TEMPERATURE ALLOWED MIN. TEMP ALLOWED MAX. TEMP MAT. PROC. CONN. + SEAL TYPE DIP STATUS MAT. FILLING FLUID SENSOR TEMP. Tmin SENSOR Tmax SENSOR MEAS. VAL. TREND ELECTR. SERIAL NO. OUTPUT FAIL MODE Endress+Hauser...
  • Page 6 HistoROM AVAIL. HistoROM CONTROL SAFETY LOCKSTATE SAFETY PASSWORD DIGITS SET DIGITS SET ACK. ALARM MODE MEASURING MODE CALIB. OFFSET SET LRV SET URV DAMPING TIME CONF. PASSWORD CURRENT OUTPUT For further information, see the Cerabar S Functional Safety Manual (SD00190P). Endress+Hauser...
  • Page 7: Graphic Representation Of Function Groups

     PROCESS VALUES pressure   OPERATING   DIAGNOSTICS (562)  SIMULATION MODE   MESSAGES   USER LIMITS   SERVICE (561)  SYSTEM 2  For further information, see the Cerabar S Functional Safety Manual (SD00190P). Endress+Hauser...
  • Page 8: Pressure Measurement

    (20 mA). Result: The measuring range is configured for 0 to +300 mbar (4.5 psi). Note! • You can also specify a customer-specific unit. See parameter description for PRESS. ENG. UNIT (Page 15). Endress+Hauser...
  • Page 9: Calibration Without Reference Pressure

    • You can also perform calibration without reference pressure by means of the QUICK SETUP menu. See Page 11 ff, Table 2: QUICK SETUP menu. • You can also specify a customer-specific unit. See parameter description for PRESS. ENG. UNIT (Page 15). Endress+Hauser...
  • Page 10: Description Of Parameters

    • Select the menu language for the operating program via the "Options" menu  "Settings" "Language" tab "Tool Language" field. Options: • Deutsch • English • Français • Italiano • Español • Nederlands • Chinese (CHS) • Japanese (JPN) Factory setting: English Endress+Hauser...
  • Page 11: Measuring Mode

    The following applies: CALIB. OFFSET = MEASURED VALUE – POS. INPUT VALUE, here: CALIB. OFFSET = 0.5 mbar (0.0075 psi) – 2.0 mbar (0.03 psi) = – 1.5 mbar (-0.0225 psi) – The current value is also corrected. Factory setting: Endress+Hauser...
  • Page 12 Recommendation: • The pressure difference between zero (set point) and the measured pressure need not be known. – POS. INPUT VALUE • The pressure difference between zero (set point) and the measured pressure is known. – CALIB. OFFSET Endress+Hauser...
  • Page 13: Calib. Offset

    To correct the MEASURED VALUE to 0.0 mbar, you must enter the value 2.2 here. (MEASURED VALUE = MEASURED VALUE – CALIB. OFFSET) – MEASURED VALUE (after entry for calib. offset) = 0.0 mbar – The current value is also corrected. Factory setting: Endress+Hauser...
  • Page 14: Measuring Mode

    The operating menu is structured according to the selected measuring mode. Note! When the measuring mode is changed, no conversion takes place. The device has to be recalibrated if the measuring mode is changed. Prerequisite: • Digital communication Options: • Pressure Factory setting • Pressure Endress+Hauser...
  • Page 15: Press. Eng. Unit

    0.0 or as per order specifications SET URV (246) Set upper range value – without reference pressure. Entry Enter pressure value for the upper current value (20 mA). Factory setting: High sensor limit ( see PRESS. SENS HILIM, Page 28) Endress+Hauser...
  • Page 16 The damping affects the speed at which all subsequent elements, such as the onsite display, measured value and current output react to a change in the pressure. Input range: 0.0 to 999.0 s Factory setting: 2.0 s or as per order specifications Endress+Hauser...
  • Page 17 • See also Page 8 ff, Section 4 "Pressure measurement". TEMP. ENG. UNIT (318) Select the unit for the temperature measured values. See also PCB TEMPERATURE (Page 25) and SENSOR TEMP. (Page 30). Options Options: • °C • °F • K • R Factory setting: °C Endress+Hauser...
  • Page 18: Menu Descriptor

    Specifies the number of places after the decimal point for the value displayed in the main Options line. See also Operating Instructions BA00412P, Section "Onsite display". Options: • Auto • x.x • x.xx • x.xxx • x.xxxx • x.xxxxx Factory setting: Auto Endress+Hauser...
  • Page 19: Display Contrast

    4...13, 4: contrast weaker (brighter), 13: contrast stronger (darker). Factory setting: DIGITS SET (840) This parameter is used to verify correct display of characters and digits on the user Display interface. If the characters and digits are correctly displayed, this parameter shows the string "0123456789.-". Endress+Hauser...
  • Page 20: Output Current

    SET MIN. CURRENT Level measuring ASSIGN CURRENT mode only 1) Display via on-site display only P01-MIDxxxxx-19-xx-xx-EN-013 Fig. 9: Group OUTPUT Table 7: (GROUP SELECTION ) OPERATING MENU OUTPUT Parameter name Description OUTPUT CURRENT (254) Displays the current current value. Display Endress+Hauser...
  • Page 21: Output Fail Mode

    • Normal: the current output assumes the value set via the OUTPUT FAIL MODE and SET MAX. ALARM parameters. • NAMUR: – Lower sensor limit undershot (E120): Current output = 3.6 mA – Upper sensor limit overshot (E115): current output assumes the value set via the SET MAX. ALARM parameter Factory setting: Normal Endress+Hauser...
  • Page 22: Set Max. Alarm

    Input range: 21...23 mA Factory setting: 22 mA SET MIN. CURRENT (343) Enter lower current limit. Entry Some switching units do not accept current values lower than 4.0 mA. Options: • 3.8 mA • 4.0 mA Factory setting: 3.8 mA Endress+Hauser...
  • Page 23: Bus Address

    (HART 5.0: range 0...15, HART 6.0: range 0...63) Factory setting: DEVICE TYPE (802) Displays the device identification number in decimal numerical format, Display here Cerabar S: 24 Prerequisite: • Cerabar S pressure transmitter DEVICE REVISION (699) Displays the device revision Display Endress+Hauser...
  • Page 24: Preamble Number

    Factory setting: MANUFACTOR ID (432) Displays the manufacturer number in a decimal numerical format. Display Here: 17 Endress+Hauser HART MESSAGE (271) Enter a message (max. 32 alphanumeric characters). Entry On command from the master, this message is sent via the HART protocol.
  • Page 25: Cust. Tag Number

    PCB TEMPERATURE (357) Displays the measured temperature of the main electronics. Display ALLOWED MIN. TEMP Displays the lower temperature limit of the main electronics. (358) Display ALLOWED MAX. TEMP Displays the upper temperature limit of the main electronics. (359) Display Endress+Hauser...
  • Page 26: Mat. Proc. Conn

    • Steel • 304 st. steel • 316 st. steel • Alloy C • Monel • Tantalum • Titanium • PTFE (Teflon) • 316L st. steel • PVC • Inconel • PVDF • ECTFE Factory setting: As per order specifications Endress+Hauser...
  • Page 27: Seal Type

    Prerequisite: • Digital communication DIAPHRAG. MAT. - For selecting and displaying the diaphragm material on the negative side. Prerequisite: • Digital communication NR OF REMOTE SE For selecting and displaying the number of diaphragm seals. Prerequisite: • Digital communication Endress+Hauser...
  • Page 28: Sensor Ser. No

    Tmin SENSOR (368) Displays the lower nominal temperature limit of the sensor. Display Tmax SENSOR (369) Displays the upper nominal temperature limit of the sensor. Display SENS H/WARE REV (487) Displays the revision number of the sensor hardware. Display e.g.: 1 Endress+Hauser...
  • Page 29 Displays the measured value In the "Pressure" measuring mode, this value corresponds to the PRESSURE parameter. Prerequisite: • Digital communication Onsite operation: • In the case of local operation, the MEASURED VALUE parameter is displayed on the top level. Endress+Hauser...
  • Page 30: Sensor Temp

     See also PRESSURE diagram. Display SENSOR TEMP. (367) Displays the temperature currently measured in the sensor. This can deviate from the Display process temperature. MEAS. VAL. TREND (378) Displays the trend of the pressure measured value. Display Possibilities: increasing, decreasing, constant Endress+Hauser...
  • Page 31: Enter Reset Code

    ENTER RESET CODE (047) Reset parameters completely or partially to factory values or delivery status. See also Operating Instructions BA00412P, Section "Factory setting (reset)". Entry Factory setting: OPERATING HOURS (409) Displays the hours of operation. This parameter cannot be reset. Display Endress+Hauser...
  • Page 32: Download Select

    /M-DAT (optional)". Prerequisite: ® • A HistoROM /M-DAT is attached to the electronic insert (HistoROM AVAIL. = Yes) Options: • Abort • HistoROM Device • Device HistoROM Factory setting: ® Abort (if HistoROM /M-DAT is attached to the electronic insert) Endress+Hauser...
  • Page 33: Alarm Status

    • The ALARM STATUS parameter shows all the messages in descending order of priority. You can scroll through all the messages present with the O or S key. Operating program • The "Status" field and the ALARM STATUS parameter show the message with the highest priority. Endress+Hauser...
  • Page 34: Ack. Alarm Mode

    1. Enter the corresponding message number for ERROR No. field. 2. Select "Alarm" or "Warning" option. Digital communication 1. Enter the corresponding message number via the ERROR No. parameter. 2. Use the SELECT ALARMTYPE parameter to select the "Alarm" or "Warning" option. Endress+Hauser...
  • Page 35: Pmin Alarm Window

    See also these Operating Instructions, Section 6.1 "Messages" Table, code E733 and Section 6.2 "Response of outputs to errors". Factory setting: Upper sensor temperature application limit +10 K (For the upper temperature application limit, see Tmax SENSOR) Endress+Hauser...
  • Page 36: Curr. Trim 20Ma

    4. Select the SYSTEM 2 group. (Menu path: (GROUP SELECTION) OPERATING MENU SERVICE) 5. Enter the current value measured with the switching unit in the "CURR. TRIM 20mA" parameter. Input range: Measured current ±0.2 mA Factory setting: 20 mA Endress+Hauser...
  • Page 37: Offset 4Ma Trim

    Displays the difference between 4 mA and the value entered for the "CURR. TRIM 4mA" Display parameter. Factory setting: OFFSET 20mA TRIM (044) Displays the difference between 20 mA and the value entered for the "CURR TRIM Display 20mA" parameter. Factory setting: Endress+Hauser...
  • Page 38: Troubleshooting

    • If the device detects a defect in the onsite display during initialization, special error messages are generated. For the error messages, see Page 45, Section 6.1.1 "Onsite display error messages". • For support and further information, please contact Endress+Hauser Service. Code...
  • Page 39 Spans for the "Height linearized" option: LEVEL MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN. 613 (W613) Warning I>Simulation is active – Simulation is switched on, i.e. the – Switch off simulation. device is not measuring at present. Endress+Hauser...
  • Page 40 – Carry out calibration again. (See B>X-VAL. of lin. table out of edit linearization table is either below also these Operating Instructions, limits. the value for HYDR. PRESS MIN. or Section 5.) MIN. LEVEL or above the value for HYDR. PRESS. MAX. or LEVEL Endress+Hauser...
  • Page 41 ( See also Page 28, parameter Warning description for Tmin SENSOR.) – Unsuitable download carried out. – Check configuration and perform download again. – Loose connection at sensor cable – Wait a short period of time and tighten the connection, or avoid loose connection. Endress+Hauser...
  • Page 42 – Pressure measured value has – Check system/pressure measured overshot the value specified for the value. Factory Pmax ALARM WINDOW – Change value for Pmax ALARM setting: parameter. WINDOW if necessary. (See also Warning Page 35, parameter description for PmaxALARM WINDOW). Endress+Hauser...
  • Page 43 TI00383P. – Main electronics defective. – Replace main electronics. 745 (W745) Warning C>Sensor data unknown – Sensor does not suit the device – Replace sensor with a suitable (electronic sensor nameplate). sensor. Device continues measuring. Endress+Hauser...
  • Page 44 Alarm B>Memory failure in signal – Electromagnetic effects are greater – Block off electromagnetic effects or processor than specifications in the technical eliminate source of disturbance. data. See Technical Information TI00383P. – Main electronics defective. – Replace main electronics. Endress+Hauser...
  • Page 45: Response Of Outputs To Errors

    122 for "Sensor not number such as 613 for "Simulation is number such as 731 for "URV user limits connected". active". exceeded". Menu path: (GROUP SELECTION ) OPERATING MENU OUTPUT Menu path: (GROUP SELECTION ) OPERATING MENU MESSAGES Endress+Hauser...
  • Page 46: Confirming Messages

    (Section 6.1). Once you have suppressed this message using the F key, the message with the next highest priority is displayed. You can use the F key to suppress each message, one after the other. The ALARM STATUS parameter continues to display all the messages present. Endress+Hauser...
  • Page 47: Appendix

    PRESS. ENG. UNIT parameter. These parameters are indicated with a "*". • For a description of the parameters, see Section 5 "Description of parameters". The exact dependency of individual parameters on one another is explained here. Endress+Hauser...
  • Page 48 There are parameters that are only displayed if other parameters are appropriately configured. For example the CUSTOMER UNIT P parameter is only displayed if the "User unit" option was selected for the PRESS. ENG. UNIT parameter. These parameters are indicated with a "*". P01-MIDxxxxx-19-xx-xx-EN-025 Endress+Hauser...
  • Page 49 PMP71 with MID) SET URV DAMPING VALUE CONF. PASSWORD There are parameters that are only displayed if other parameters are 2) Display via appropriately configured. HART handheld terminal only These parameters are indicated with a "*". See Safety Manual SD00190P. P01-MIDxxxxx-19-xx-xx-en-032 Endress+Hauser...
  • Page 50 NR OF REMOTE SEAL 4 TH VALUE FILL FLUID 2) Display via FieldCare and HART handheld terminal only There are parameters that are only displayed if other parameters are appropriately configured. These parameters are indicated with a "*". P01-MIDxxxxx-19-xx-xx-en-033 Endress+Hauser...
  • Page 51 Pressure Level Tank content Current SIM. PRESSURE SIM. LEVEL SIM. TANK CONT. SIM. CURRENT SIM. ERROR NO. There are parameters that are only displayed if other parameters are appropriately configured. These parameters are indicated with a "*". P01-PMx7xxxx-19-xx-xx-xx-044 Endress+Hauser...
  • Page 52 Appendix Cerabar S 4 to 20mA HART Endress+Hauser...
  • Page 53 Cerabar S 4 to 20mA HART Appendix Endress+Hauser...
  • Page 54: Index

    POS. INPUT VALUE (563) ..... 11, 13 ENTER RESET CODE (047) ..... . . 31 Endress+Hauser...
  • Page 55 Tmin ALARM WINDOW (334) ..... 35 Tmin SENSOR (368) ......28 Endress+Hauser...
  • Page 56 www.endress.com/worldwide BA00413P/00/EN/03.14 71270370 CCS/FM+SGML9.0 71270370...