7. Using an M5 allen wrench, remove the six screws holding on the spectrometer header
assembly (Figure B-51).
Figure B-51 Spectrometer Header Assembly Screws
8. Remove the spectrometer header assembly (Figure B-52).
B-36
VS Series Component Leak Detector
Wear non-powdered, ESD-safe Nitride or equivalent gloves (not
NOTE
included in kit) to prevent skin oils from getting on vacuum
surfaces. Ensure that personnel are properly grounded before
proceeding.
Figure B-52 Spectrometer Assembly Removal