Setting a Reference for Smart Remote Mode

1.
Turn on the test tool and the remote. Let the DSP-FTA4XX units warm up
for five minutes. Set the remote end configuration to Smart Remote Mode.
2.
Choose four known-good SC/ST patch cords of fiber type matching the
cables to be tested, e.g., 50 µm multimode; 62.5 mulitmode, or
singlemode. Clean all fiber endfaces.
If the fiber to be tested does not use ST adapters, select patch cords
with SC adapters on one end and the appropriate adapters on the
other end. Mate the cables with the appropriate adapters.
3.
Make the connections shown in Figure 5.
Turn the rotary switch to SPECIAL FUNCTIONS. Select Set Fiber
4.
Reference; then press E.