Cutout Reset Mode

The cutout reset mode determines whether the cutout resets automatically when
the bath temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,

Press "SET" to access the parameter setting. Normally the cutout is set for
manual mode.
= 
To change to manual reset mode press "UP" and then "SET".
=

Stir Mode Select

This parameter along with the Stir set-point allows the user to set the
temperature at which the stir motor is activated. This setting is generally used
when salt is used for the bath medium. For example, you can set the mode to
"auto" and the temperature to 200 °C. This allows the stir motor to shut off and
turn on only when the salt is a liquid (>200 °C) preventing the stir motor from
overheating and or being damaged.

Press "SET" to access the parameter setting.
=
To change the setting to always on press the "UP" or "DOWN" buttons and then
"SET". When set to "Str=ON" the stir motor comes on regardless of the
temperature set in the stir set-point parameter.
Stir Mode Select defaults to "Auto" each time the power of the bath
is cycled off and back on. Therefore, this parameter has to be set
each time the bath is powered on if "Str-ON" is the desired mode of
operation.
cutout reset mode parameter
cutout set for automatic reset
cutout set for manual reset
The parameter is indicated by, Stir mode selection
parameter
Stir motor is set for automatic activation at the stir
set-point temperature.