DO use a 30 mA ground fault interrupt device. This unit
contains a liquid. A ground fault device is advised in case
liquid is present in the electrical system and could cause an
electrical shock.
Always replace the power cord with an approved cord of the
correct rating and type. If you have questions, contact an
Authorized Service Center (see the How to Contact Fluke
Calibration section).
High voltage is used in the operation of this equipment.
Severe injury or death may result if personnel fail to observe
the safety precautions. Before working inside the
equipment, turn off the power and disconnect the power
cord.
Keep all combustible materials away from the bath when
using salt. Operate the bath on a heatproof surface such as
concrete. Provide a means of safety for containing any spill,
which may occur.
BATH SALT
Salt used in this unit may produce noxious or toxic fumes
under certain circumstances. Consult the fluid
manufacturer's MSDS (Material Safety Data Sheet). Proper
ventilation and safety precautions must be observed.
The unit is equipped with a soft cutout (user settable
firmware) and a hard cutout (set at the factory). Check the
flash point, boiling point, or other fluid characteristic
applicable to the circumstances of the unit operation.
Ensure that the soft cutout is adjusted to the fluid
characteristics of the application. Failing to set the cutout to
the limits of the bath fluid can result in fire hazards and
personal injury.