Fluke 6050H User Manual page 20

Calibration bath
Hide thumbs Also See for 6050H:
Table of Contents

Advertisement

6050H
User's Guide
Press "SET" to select this memory and access the set-point value.
Press "UP" or "DOWN" to change the set-point value.
Press "SET" to accept the new value and display the vernier value. The bath
begins heating or cooling to the new set-point.
Press "EXIT" and the bath temperature will be displayed again.
The bath heats or cools until it reaches the new set-point temperature. Set the
heater switch to position "MED" or "HIGH" to allow the bath to more quickly reach
a higher temperature. The higher heater settings may be necessary to reach
higher temperatures and control at high temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set to
prevent this from happening. See the Cutout section.
If operating the bath below 45 °C set the COOLING power switch to ON. The
cooling temperature may require adjustment to provide the proper amount of
cooling. See the Cooling section.
To obtain optimum control stability adjust the proportional band as discussed in
the Proportional Band section.
12
Access set-point value
  

Increment display
  

Store new set-point, access vernier
 

Return to the temperature display
  

Current value of set-point 1, 25.00 °C
New set-point value
Current vernier value
Bath temperature display