P377 Initial Zero
The initial zero is the reference zero to which all subsequent operator initiated zero
calibrations are compared in determining whether they have deviated beyond the zero
limit (P370).
Entry:
0 = idle
1 = initial zero
Note:
Refer to
P388 Initial Span
The initial span is the reference to which all subsequent span calibrations are compared
in determining whether they have deviated beyond 12.5% of the initial span.
Entry:
0 = idle
1 = initial span
Note:
Refer to R

Linearization Parameters (P390 - P396)

Note:
These parameters are used to compensate for non-linear response of the
weighing system to the BW100. Refer to Recalibration/Linearization for details and
example on the use of these parameters.
P390 Linearization
Enables or disables the linearization function.
Entry:
0 = OFF
1 = ON
P391 Linearizer, Point 1
Enter the load, in units of P017, for point 1
P392 Compensation, Point 1
Enter the calculated compensation, in percent, for compensation point 1
7ML19985DJ01
Milltronics BW100 - INSTRUCTION MANUAL
Recalibration/Initial Zero
ecalibration/Initial Span
on page 33 for use of this function.
on page 36 for use of this function.