Programme Chart - Miele G 7883 Operating Instructions Manual

Laboratory glassware washer
Table of Contents

Advertisement

Programme chart

Programme
Custom
Inorganica
Organica
Standard
Universal
Intensive
Plastic
Drain
Rinse
varioTD
Special
CW = Cold water, HW = Hot water, DI = De-ionized water, °C = Temperature, min = Holding time,
* The dispensing of rinsing agent with DOS 2 must be activated by a Miele service technician
72
Freely programmable programme (programmed by a Miele service technician).
For the removal of inorganic residues;
general analysis and water analysis, suitable for low to medium soiling levels and
for medium to high quality rinse requirements.
For the removal of organic residues such as oil, grease, wax and agar. Suitable
for preparation and analytical applications with medium to high levels of soiling
and medium quality rinse requirements. Requires liquid cleaning agent: Hot,
cold and DI water connections recommended.
Suitable for a variety of soil types which do not denaturate or break down in
alkalis. Suitable for preparation and analytical applications with low levels of
soiling and average rinse requirements.
For the removal of organic residues such as protein, oil, certain types of grease,
and for certain inorganic residues (pH 7 water soluble metal salts). Suitable for
preparation and analytical applications with low to medium levels of soiling and
medium quality rinse requirements.
For the removal of organic residues such as protein, oil, cell and tissue culture,
oils and certain types of grease (pH 7 water soluble metal salts). Suitable for
preparation and analytical applications with medium to high levels of soiling and
medium to high quality rinse requirements.
For temperature sensitive laboratory equipment e.g. plastic bottles (temperature
resistant to at least 55°C). Suitable for preparation and analytical applications
with low to medium levels of soiling and medium quality rinse requirements.
To drain soiled water from the machine, e.g. after a programme has been
cancelled. If applicable, observe regulations on containing the spread of
epidemics - see "To cancel a programme".
For rinsing heavily soiled items (e.g. for removing soiling or disinfectant residues,
or to prevent soiling drying on to laboratory glassware and utensils when a
complete programme does not yet need to be run).
Complies with general hygiene regulations. 90°C (+ 5 °C, - 0 °C) programme with
a 5-minute holding time.
For cleaning and thermal disinfection at 93°C with a 10 minute holding time, area
of effectiveness A/B range
When to use

Hide quick links: