Miele G 7883 Operating Instructions Manual page 67

Laboratory glassware washer
Table of Contents

Advertisement

Levelling and securing the machine
To ensure stability, the machine must
first be aligned and then screwed to the
worktop.
^ Open the door.
^ Secure the machine to the front edge
of the worktop using screws through
the holes on the left and right hand
sides of the front trim.
^ Do not use silicone sealant to seal
the gaps between the machine and
any neighbouring units, as this would
hinder ventilation to the circulation
pump.
Cover plate (protects the worktop)
If required, a special cover plate can
be ordered from the Miele Service
Department.
This cover plate protects the underside
of the worktop against steam rising
from the machine.
Decor frame and decor panel
(optional)
The machine front can be fitted with a
decor panel to match kitchen units.
This must be done by a Miele Service
technician. Contact Miele for details.
Decor panel dimensions:
Door outer panel H 442 / W 586 mm
Installation flap
Installation
H 117 / W 586 mm