Essential maintenance

Maintenance

This lab washer should be serviced by
a Miele service technician in
accordance with local and national
safety regulations after every 1000
operating hours, or annually.
A service will cover the following:
– Electrical safety tests
– Door mechanism and door seal
– Any screw connections and
connectors in the wash cabinet
– Water inlet and drainage
– Internal and external dispensing
systems
– Spray arms
– Filter combination
– Sump, including drain pump and
non-return valve
– All baskets, inserts and modules
and where applicable:
– Steam condenser
– Any printer connected to the
machine.
46
The following operational tests will be
carried out within the framework of the
maintenance:
– A programme will be run as a test
run
– Thermo electrical measurements will
be taken
– Seals will be tested for water
tightness
– All relevant measuring systems will
be safety tested including fault
indications.