Table 2 9103 Controller Communications Commands - Fluke 9103 User Manual

Dry-well
Hide thumbs Also See for 9103:
Table of Contents

Advertisement

9103 Dry-Well
User's Guide

Table 2 9103 Controller Communications Commands

Command Description
Display Temperature
Read current set-point
Set current set-point to n
Read temperature
Read temperature units
Set temperature units:
Set temperature units to Celsius
Set temperature units to
Fahrenheit
Read scan mode
Set scan mode
Read scan rate
Set scan rate
Read hold
Secondary Menu
Read proportional band setting
Set proportional band to n
Read heater power
(duty cycle)
Configuration Menu
Operating Parameters Menu
Read high limit
Set high limit
Serial Interface Menu
Read serial sample setting
Set serial sampling setting to n
seconds
Set serial duplex mode:
Set serial duplex mode to full
Set serial duplex mode to half
Set serial linefeed mode:
Set serial linefeed mode to on
Set serial linefeed mode to off
Calibration Menu
34
info@Fluke-Direct.com
Command
Command
Format
Example
s[etpoint]
s
s[etpoint]=n
s=200.0
t[emperature]
t
u[nits]
u
u[nits]=c/f
u[nits]=c
u=c
u[nits]=f
u=f
sc[an]
sc
sc[an]=on/off
sc=on
sr[ate]
sr
sr[ate]=n
sr=1.1
ho[ld]
ho
pr[opband]
pr
pr[opband]=n
pr=8.83
po[wer]
po
hl
hl
hl=n
hl=90
sa[mple]
sa
sa[mple]=n
sa=0
du[plex]=f[ull]/h[alf]
du[plex]=f[ull]
du=f
du[plex]=h[alf]
du=h
lf[eed]=on/of[f]
lf[eed]=on
lf=on
lf[eed]=of[f]
lf=of
Fluke-Direct
Returned
Returned
Example
set: 999.9 {C or F}
set: 100.00 C
t: 999.9 {C or F}
t: 55.6 C
u: x
u: C
scan: {ON or OFF}
scan:ON
srat: 99.9 {C or F}/min
srat:12.4C/min
hold: open/closed, 99.9
hold: open, 30.5
{C or F}
C
pb: 999.9
pb: 15.9
po: 999.9
po: 1.0
hl:999
hl:126
sa: 9
sa: 1
.com
Acceptable
Values
Instrument
Range
C or F
ON or OFF
.1 to 99.9
Depends on
Configuration
0–126
0 to 999
FULL or HALF
ON or OFF