Service and maintenance

6.1

Maintenance of analysis module

6.1.1

Removal of analysis module

Procedure
Explosion hazard
 If the gas chromatograph is mounted in a hazardous area, observe the explosion
 Before opening the housing in the hazardous area, switch off the power supply and then
 Following the waiting time:
1. Loosen the 1.5-mm hexagon socket-head bolt
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
WARNING
protection regulations.
always wait for at least 65 minutes.
– When the surface temperatures inside the gas chromatograph have cooled down to
a temperature corresponding to temperature class T4.
– When the electric charge on the internal components has dropped to a voltage
permissible for explosion protection.
turns.
in the top cover
6
by two complete
77