Note
Types of solenoid valves
The membrane valve functions are controlled by solenoid valves. Two types of solenoid
valves are used: open when deenergized and closed when deenergized.
Closed when deenergized: MV2 - Sample, MV4 - Injection
Open when deenergized: MV1 - Flush loop, MV3 - Back flush
Open when deenergized: MV1 – Flush loop_1, MV5 – Flush loop_2
Result
The result of the measurement is e.g. a chromatogram with time switching points on detector
TCD 2. This chromatogram shows the difference in the two procedures (live and eLive)
between the switching points t
t
1
Figure 3-12
3.6.5.1

Preparation of measurement

Note
Time switching points
The time switching points shown on the following pages all depend on the type of
application. Only the time switching point t
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
t
t
2
3
Example: chromatogram with time switching points
and t
.
4
6
for the injection is always zero.
6
Description
3.6 Functions