Danger of explosion through discharged gases
In the event of a fault, all gases supplied to the chromatograph (also samples and hydrogen
as carrier gas) can be discharged via the breathing connector or a connected line. In such a
case, there is a danger to personnel, equipment and the environment!
For this reason, take appropriate measures to prevent this.
Never connect the breathing connector to the exhaust gas system.
4.4.9
Power switch and fuse
The gas chromatograph does not have a power switch. Therefore install an external power
switch with a switching capacity of min. 2.5 A and a slow-blow fuse with a rating of 2.5 A.
Note
Fit the power switch such that it is readily accessible during operation.
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
DANGER
Mounting
4.4 Mounting and connecting
63