Mounting
4.4 Mounting and connecting
4.4.8

Breathing device

The breathing device is used for pressure compensation during operation. Pressure
compensation is required due to temperature variations during operation
Figure 4-3
Danger
Danger of explosion through discharged gases
In the event of a fault, all gases supplied to the chromatograph (also samples and hydrogen
as carrier gas) can be discharged via the breathing connector or a connected line. In such a
case, there is a danger to personnel, equipment and the environment!
 For this reason, take appropriate measures to prevent this.
 Never connect the breathing connector to the exhaust gas system.
Do not close off the breathing device
Ensure that the breathing device is not closed during operation, otherwise:
 Pressure compensation is no longer carried out.
 Ex protection is no longer guaranteed.
 Housing protection IP65 is only guaranteed if the pipeline connected to
62