Download Print this page

Troubleshooting; Indicator Lights - Invacare HomeFill II IOH200AW User Manual

Compressor
Hide thumbs Also See for HomeFill II IOH200AW:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

6 Troubleshooting

6.1 Indicator Lights

All indicators are for low priority conditions that require operator awareness only.
INDICATOR LIGHT COLOR
None
O
Below Normal -
2
YELLOW
Filling - GREEN
1145804-E
COMPRESSOR STATUS
Compressor is Off.
Compressor is On and warming up
(0-3 minutes after power is On).
O
from concentrator is < 90% (any time after
2
three minute warmup).
Compressor is On and cylinder is filling. O
from concentrator is >90% (any time after
three minute warmup).
EFFECT
O
flowing to user if concentrator is On.
2
Cylinder filling not started.
O
flowing to user if concentrator is On.
2
Cylinder filling stops, will restart when O
> 90%.
O
flowing to user if concentrator is On.
2
Cylinder is filling
2
O
flowing to user if concentrator is On.
2
Troubleshooting
2
23

Advertisement

loading