Invacare® HomeFill® II

2.5 EMC Information

Medical Electrical Equipment needs to be installed and used according
to the EMC information in this manual.
This equipment has been tested and found to comply with EMC
limits specified by IEC/EN 60601-1-2 for Type BF equipment.
These limits are determined to provide reasonable protection
against electromagnetic interference in a typical home healthcare
environment. This equipment is not intended for use in a professional
healthcare environment.
Portable and mobile RF communications equipment can affect the
operation of this equipment. The compressor may be stacked on top
on an Invacare HomeFill compatible oxygen concentrator during
setup. The compressor operation should be observed to verify
normal cylinder filling in either situation.
Other devices may experience interference from even the low levels
of electromagnetic emissions permitted by the above standard. To
determine if the emissions from the compressor are causing the
interference, turn the compressor Off. If the interference with the
other device operation stops, then the compressor is causing the
interference. In such rare cases, interference may be reduced or
corrected by one of the following measures:
Reposition, relocate, or increase the separation between the
devices.
Connect either line powered device to a different electrical
power circuit.
Contact Invacare for any additional EMC information or EMC tables
for the device environment.
12
1145804-E