Appendix A
Device-Specific Information
E Series User Manual
Differential Signals
To connect differential signals, determine the type of signal source you are
using: a floating signal source or a ground-referenced signal source. Refer
to the
Differential Connection Considerations
Input Signals
sections of Chapter 2,
AI signals.
To measure a floating signal source, move the switch to the FS position. To
measure a ground-referenced signal source, move the switch to the GS
position. Figure A-57 shows the source type switch locations on the front
panel of the BNC DAQPads.
FS = Floating Source
GS = Grounded Source
USER 2
PWR
COM
1
3
5
7
9 11 13 15 17 19 21 23 25 27 29
2
4
6
8 10 12 14 16 18 20 22 24 26 28 30
USER 1
FS GS
Figure A-57. BNC DAQPad Front Panel
Figure A-58 shows the analog input circuitry on BNC DAQPads. When the
switch is in the FS position, AI x – is grounded through a 0.1 μF capacitor
in parallel with a 5 kΩ resistor.
Figure A-58. Analog Input Circuitry
Analog
CTR 0
EXT
DAC 0
OUT
REF
OUT
PFI 0/
DAC 1
TRIG 1
OUT
FS GS
GS
FS
FLOATING SOURCE
GND REF SOURCE
AI x +
AI x –
GS
FS
0.1 µF
5 kΩ
AI GND
A-72
and
Connecting Analog
Input, for more information on
Source Type
Switch
ACH 6
ACH 4
ACH 2
ACH 0
FS GS
FS GS
FS GS
FS GS
ACH 7
ACH 5
ACH 3
ACH 1
FS GS
FS GS
FS GS
Source Type
Switch
ni.com